Dose-Rate

Dose Rate Measurements

Typical dose measurements obtained during the acceptance test at installation time are depicted below. The total charge after 360 seconds was repeatable within 1%.

Verifying collimator output factors: Output factors were obtained with average readings using MOSFET diode at the center of the acrylic phantom. The output factors for 18 mm, 14 mm and 8 mm were also measured with the ion chamber as indicated.

Size MFR Factors Meausured with Ion Chamber Gafchromic Measurement Dose Rate
cGy/min
18 mm 1.000 1.00 1.00 367
14 mm 0.95 0.95 0.92 338
8 mm 0.85 0.68 N/A 0.87 320
4 mm 0.75 0.32 N/A 0.73 268
Closed 0.01 0.002 N/A
 